INNOVATION  
Year : 2016  |  Volume : 9  |  Issue : 3  |  Page : 201-203
Finger-shaped red light emitting diode to ascertain the depth of periungual wart


Department of Dermatology, Christian Medical College, Vellore, India

Click here for correspondence address and email

Date of Web Publication7-Oct-2016
 

   Abstract 

Management of periungual wart is a great challenge, especially when there is subungual extension. The major cause of recurrence of wart is improper clinical assessment of its extent and not directing therapy against the entire wart. This difficulty of ascertaining its extent could be overcome with this finger-shaped red light emitting diode device. Red light in the device penetrates the thick palmar skin and dark constitutive skin colour due to its longer wavelength.

Keywords: Periungual, red, wart

How to cite this article:
Nirmal B. Finger-shaped red light emitting diode to ascertain the depth of periungual wart. J Cutan Aesthet Surg 2016;9:201-3

How to cite this URL:
Nirmal B. Finger-shaped red light emitting diode to ascertain the depth of periungual wart. J Cutan Aesthet Surg [serial online] 2016 [cited 2019 Nov 14];9:201-3. Available from: http://www.jcasonline.com/text.asp?2016/9/3/201/191655



   Introduction Top


Warts are the most common benign nail tumours. Warts around the nail unit are commonly caused by human papillomavirus – 1, 2 and 4. Periungual warts originate from the skin possessing granular layer including hyponychium, proximal and lateral nail folds. They can invade the nail bed and compress the matrix.[1] The current management options for periungual warts include topical, intra-lesional and systemic drugs, lasers and surgical procedures. The choice of a procedure in a patient depends on skill of the treating physician, location and extent of wart, number of warts, duration of warts, age and immunologic status of the patient.[2] Nevertheless, periungual warts are known to recur and one of the most common causes of recurrence is inability of treating physician to determine the extent of it. We designed a simple device to ascertain the extent of periungual wart.


   Device Specifications Top


The device is handy with dimensions of 8.5 cm × 5.5 cm × 2.5 cm [Figure 1]a. Power source for the light emitting diode (LED) device is a 9V battery with a switch adjacent, both fixed at the upper part of the device [Figure 1]b. A row of nine 3.3V, 20 mA red colour LEDs is fixed on a plastic board and is enclosed in a black opaque cover [Figure 1]c. The opaque cover transmits light only to the bottom and there is no dispersion of light through the sides [Figure 1]d. The total cost incurred to design the device is 200 INR [Table 1].
Figure 1: (a) Finger-shaped red light emitting diode device; side view: battery on top, (b) top view: 9V battery power source with a switch adjacent, (c) bottom view: row of nine light emitting diodes on a plastic board, (d) opaque black cover allows light to pass only through bottom

Click here to view
Table 1: Cost of making the device

Click here to view



   Principle Top


Visible spectrum has a colour range from violet to red, with a wavelength ranging from 400 to 700 nm. The penetration depths of these colours depend on their wavelength.[3] The light with longer wavelength penetrates deeper. Blue light has a wavelength range of 400–470 nm and can penetrate only up to 1 mm. Red has the longest wavelength (630–700 nm) in the visible spectrum and can penetrate 1–6 mm [Figure 2].
Figure 2: Wavelength of light and depth of penetration : Across the visible spectrum from violet to red, the light penetrates deeper with higher wavelengths

Click here to view



   Discussion Top


The LED device fits snugly when placed with its bottom on the finger [Figure 3]. As there is no dispersion of light through the sides, when illuminated, the red light passes onto the finger. Wart stands out contrasting as black against a bright red background making it easy for the treating physician to appreciate its extent [Figure 4]. Trans-illumination with white torch light is not sufficient. Limitations of white light are it is ineffective in thicker palmoplantar skin and darker constitutional skin colour.[4] These limitations could be overcome with this device due to deeper penetrating red light.
Figure 3: Light emitting diode device fits snugly when placed with its bottom on the finger

Click here to view
Figure 4: Periungual wart stands out black against a bright red background

Click here to view


Therapies for periungual wart are targeted only to that portion visible to the naked eye. Subungual extension of wart can be missed with the naked eye examination. For subungual extension, most treatment modalities require a partial nail avulsion before targeting therapy except for those having immunomodulatory action like bleomycin. This device addresses this problem of assessment and delineates the complete extent of the periungual wart with distinct advantages [Table 2].
Table 2: Advantages of the device

Click here to view


The red LED device can also be used in injecting sclerotherapy agents for varicose veins and drawing blood in individuals with clinically invisible veins. Knowing the extent of periungual wart is required for successful management of the condition which can be achieved with this LED device.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Tosti A, Piraccini BM. Warts of the nail unit: Surgical and nonsurgical approaches. Dermatol Surg 2001;27:235-9.  Back to cited text no. 1
[PUBMED]    
2.
Herschthal J, McLeod MP, Zaiac M. Management of ungual warts. Dermatol Ther 2012;25:545-50.  Back to cited text no. 2
[PUBMED]    
3.
Barolet D. Light-emitting diodes (LEDs) in dermatology. Semin Cutan Med Surg 2008;27:227-38.  Back to cited text no. 3
[PUBMED]    
4.
Ashique KT, Kaliyadan F. Transillumination: A simple tool to assess subungual extension in periungual warts. Indian Dermatol Online J 2013;4:131-2.  Back to cited text no. 4
[PUBMED]  Medknow Journal  

Top
Correspondence Address:
Dr. Balakrishnan Nirmal
Department of Dermatology, OPD # 220, Room # 5, Christian Medical College, Vellore - 632 004
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0974-2077.191655

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
  
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Device Specifica...
   Principle
   Discussion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1382    
    Printed16    
    Emailed0    
    PDF Downloaded74    
    Comments [Add]    

Recommend this journal