Cutaneous vesicles caused by transcutaneous gas-monitoring sensor


Chief Consultant Anesthesiologist, Wockhardt Hospitals, Bangalore, Karnataka - 560 076, India

Click here for correspondence address and email

Date of Web Publication1-Jul-2009
 

How to cite this article:
Chakravarthy M, Narayan S, Govindarajan R, Rajeev S, Jawali V. Cutaneous vesicles caused by transcutaneous gas-monitoring sensor. J Cutan Aesthet Surg 2009;2:49-50

How to cite this URL:
Chakravarthy M, Narayan S, Govindarajan R, Rajeev S, Jawali V. Cutaneous vesicles caused by transcutaneous gas-monitoring sensor. J Cutan Aesthet Surg [serial online] 2009 [cited 2019 Apr 18];2:49-50. Available from: http://www.jcasonline.com/text.asp?2009/2/1/49/53105


Sir,

Transcutaneous gas monitoring is a noninvasive technique of measurement of oxygen and carbon dioxide tension across the skin. Several studies have validated the accuracy of the values obtained from this monitor.[1],[2] Transcutaneous measurements of partial pressure of oxygen and partial pressure of carbon dioxide are based on the principle that a heating element in the electrode elevates the temperature in the underlying tissue. This increases the capillary blood flow, and the partial pressure of oxygen and carbon dioxide, thereby rendering the skin permeable to gas diffusion. The method has been claimed to be safe and free from side-effects. During a study for evaluation of TINA TCPM4 (Radiometer, Copenhagen, Denmark) for transcutaneous carbon dioxide tension and transcutaneous oxygen tension monitoring, two patients developed blister formation due to possible thermal injury. This case report documents the occurrence of such blister formation and the need for awareness amongst physicians about such a side-effect.

A total of 48 patients who had cardiac surgery during April-July 2008 in our center, underwent transcutaneous gas tension monitoring measured with TINA TCPM4 (Radiometer, Copenhagen, Denmark). During the monitoring, two patients were noticed during routine clinical examination, to have developed vesicles at the site/s of use of the sensor for the equipment on the chest [Figure 1]a and b. There was no previous history of any allergy or atopy in the patients. The first patient developed the vesicle after 3 hours and the second patient after 4 hours of application. In both the patients, the lesions were asymptomatic. There was no history of itching or burning. Dermatological examination revealed tense vesicles on a non-inflammatory base. The lesions subsided spontaneously in 24 hours.

The concept of transcutaneous measurement of respiratory gases looks promising because of its noninvasive nature. The manufacturers, in their product brochure mention the potential for the occurrence of burn injuries at the site of sensor application and recommend the use of sensor at 44C to minimize the potential risk of burn injury and pressure-induced necrosis. In two of our patients, the application of probes led to blister formation despite adhering to these instructions. The possibility of contact dermatitis was ruled out by the non-inflammatory nature, absence of itching and the occurrence of blisters within 4 hours of application of the probes. Possible mechanisms of blister formation in the patients were thermal injury caused by the increased local skin temperature and the negative pressure within the well of the sensor while the gas expelled via the skin is sampled.

In the light of our experience, we recommend that clinicians wishing to use the transcutaneous measurements may need to restrict the usage time of sensor at any site for less than 3 hours and at temperatures less than 44C. The sensors need to be applied on aesthetically less significant areas. Further, the possibility of developing such vesicles may be informed to patients during informed consent.

 
   References Top

1.Johnson DC, Batool S, Dalbec R. Transcutaneous carbon dioxide pressure monitoring in a specialized weaning unit. Respir Care 2008;53:1042-7.  Back to cited text no. 1  [PUBMED]  [FULLTEXT]
2.Fanelli G, Baciarello M, Squicciarini G, Malagutti G, Zasa M, Casati A. Transcutaneous carbon dioxide monitoring in spontaneously breathing, nonintubated patients in the early postoperative period. Minerva Anestesiol 2008;74:375-80.  Back to cited text no. 2  [PUBMED]  [FULLTEXT]

Top
Correspondence Address:
Murali Chakravarthy
Chief Consultant Anesthesiologist, Wockhardt Hospitals, Bangalore, Karnataka - 560 076
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0974-2077.53105

Rights and Permissions


    Figures

  [Figure 1]

This article has been cited by
1 Weaning Mechanical Ventilation After Off-Pump Coronary Artery Bypass Graft Procedures Directed by Noninvasive Gas Measurements
Murali Chakravarthy, Sandeep Narayan, Raghav Govindarajan, Vivek Jawali, Subramanyam Rajeev
Journal of Cardiothoracic and Vascular Anesthesia. 2010; 24(3): 451
[VIEW] | [DOI]



 

Top
 
  Search
 
  
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Email Alert *
    Add to My List *
* Registration required (free)  


    References
    Article Figures

 Article Access Statistics
    Viewed1835    
    Printed116    
    Emailed0    
    PDF Downloaded246    
    Comments [Add]    
    Cited by others 1    

Recommend this journal